Table 2.
Used failure types and their defining criteria.
| Failure type | Criteria |
|---|---|
| 1 | Visible yielding and dislocation/tilt of the tibia plateau of more than 2 mm. This criterion can only be checked in the unloaded condition and addresses the unwanted loss of correction after surgery. |
|
| |
| 2 | Visible collapse of lateral cortex. Small hairline cracks are not considered as failure. |
|
| |
| 3 | Maximal displacement range of more than 0.5 mm within one hysteresis loop in the case of cyclic testing only. |
|
| |
| 4 | Cracks of the screws of more than 1 mm. |