. 2004 May;42(5):2094–2100. doi: 10.1128/JCM.42.5.2094-2100.2004

TABLE 1.

Precisions of the Roche and Artus assays for detection of SARS CoVa

Nominal concn (no. of copies/ml) Roche assay
Artus assay
Mean measured concn (no. of copies/ml [range]) Mean Ct SD % CV Mean measured concn (no. of copies/ml [range]) Mean Ct SD % CV
106 4,974,250 (3,686,000-5,992,000) 26.38 0.22 0.82 3,401,750 (2,531,000-4,409,000) 25.52 0.21 0.83
105 514,925 (340,000-689,000) 29.58 0.25 0.84 301,950 (222,800-365,500) 28.86 0.19 0.66
104 76,688 (52,080-107,720) 32.28 0.37 1.15 30,684 (23,740-37,480) 32.00 0.18 0.55
103 22,057 (14,586-28,540) 33.98 0.15 0.45 3,716 (2,822-5,763) 34.92 0.25 0.71
102 13,661 (8,584-20,520) 34.73 0.22 0.62 439 (256-766) 37.95 0.57 1.49
101 17,460 34.33 220 (85-165) 39.14 0.88 2.24
a

External standard RNAs (a gamma-irradiated, purified Vero cell culture supernatant was used as a full-virus quantification standard) were tested at six different concentrations. Each concentration was analyzed in eight PCR runs. Ct represents the PCR cycle at which the probe-specific fluorescent signal can be detected against the background.