Skip to main content
. 2014 May 29;9:123. doi: 10.1186/1748-717X-9-123

Table 2.

RPC and VPP average percentage of pixels passing the gamma criteria at 3%/3 mm, 5%/5 mm and 7%/4 mm for 12 institutions of the EORTC for IMRT complex dosimetry checks

 
RPC
 
 
VPP
Institution number 3%/3 mm (%) RPC PH 5%/5 mm (%) RPC PH 7%/4 mm (%) RPC PH Number of verified fields Measurement device 3%/3 mm (%) [SD] INST PH 5%/5 mm (%) [SD] INST PH 7%/4 mm (%) [SD] INST PH
1
89.3
99.9
99.8
Combined
Delta4
100.0 [-]
100.0 [-]
100.0 [-]
2
62.9
82.0
91.2
Combined
PTW 2D-array
93.2 [-]
99.0 [-]
99.4 [-]
3
78.1
99.9
97.8
Combined
PTW 2D-array
90.8 [0.9]
100 [0.2]
100 [0.3]
4
84.7
98.2
99.5
Combineda
PTW 2D-array
84.1 [1.9]
95.7 [0.6]
96.4 [0.7]
5
80.1
98.2
99.3
Combineda
Varian Portal
96.5 [3.3]
99.8 [0.3]
100.0 [0.0]
6
96.4
99.8
99.9
9
Varian Portal
96.6 [2.2]
99.4 [1.3]
100.0 [0.0]
7
96.5
99.8
100.0
Combined
PTW 2D-array
85 [1.3]
96.6 [0.6]
97.3 [0.5]
8
89.1
96.9
98.1
Combined
EDR2-Film
89.1 [-]
99.3 [-]
99.9 [-]
9
73.5
91.9
95.5
Combined
Delta4
99.5 [-]
100.0 [-]
100.0 [-]
10
79.6
94.1
98.8
Combined
Delta4
98.4 [-]
100.0 [-]
100.0 [-]
11
89.9
98.9
99.7
Combined
EDR2-Film
92.2 [-]
97.0 [-]
100.0 [-]
12
72.5
93.0
96.8
Combined
Delta4
97.1 [-]
100.0 [-]
100.0 [-]
CDC
 
 
 
 
 
 
 
 
Pass at 80% pixel level
7
12
12
 
 
12
12
12
Pass at 90% pixel level
2
11
12
 
 
9
12
12
Pass at 95% pixel level 2 8 11     6 12 12

Abbreviations: RPC: Radiological Physics Center; VPP: Virtual Phantom Project; SD: Standard Deviation; CDC, complex dosimetry checks. Combined: A combined measurement of all fields is made a: measured twice.