Published in final edited form as: Ophthalmology. 2014 Feb 15;121(6):1194–1202. doi: 10.1016/j.ophtha.2013.12.027

Table 6.

Reproducibility of Anterior Segment Biometric Parameter Measurements in a Random Subset of 20 Eyes

Parameters ICC Mean Difference 95% CI
PD 0.908 4.426 0.155 0.777, 0.964
CCT (ìm) 0.979 642.79 1.34 0.945, 0.992
IT750 (mm) 0.921 0.494 0.001 0.805, 0.969
IT2000 (mm) 0.980 0.495 0.005 0.949, 0.992
ITM (mm) 0.861 0.647 0.010 0.673, 0.945
ICURV(mm) 0.885 0.15 0.01 0.724, 0.955
TISA500 (mm2) 0.977 0.125 0.001 0.942, 0.991
TISA750 (mm2) 0.979 0.244 0.002 0.947, 0.992

PD = pupil diameter, CCT = central corneal thickness, IT750 = iris thickness measured at 750 μm from the scleral spur, IT2000 = iris thickness measured at 2000 μm from the scleral spur, ITM = maximal iris thickness, ICURV = iris curvature, TISA500 = trabecular–iris space area at 500 μm from the scleral spur, TISA750 = trabecular–iris space area at 750 μm from the scleral spur, ICC = intraclass correlation coefficient, CI = confidence interval