Published in final edited form as: IEEE Trans Ultrason Ferroelectr Freq Control. 2012 Jul;59(7):1602–1608. doi: 10.1109/TUFFC.2012.2360

Table II.

Tissue displacement width, tracking beam width and the ratio of displacement width to tracking width (Disp, Track, Displ:Track, respectively) for the 1.5D ARFI and the conventional 1D array. The 1.5D ARFI array has a 9 fold increase in Dy and a 1.2 fold increase in Dx as compared to the conventional 1D array. For comparison, the ratio of pushing beam width to tracking beam width is also included.

Transducer Direction Displ (mm) Track (mm) Displ: Track (Di) Push:Track
Conventional 1D Lateral 6.4 0.25 25.9 1.5
Elevation 10.3 5.4 1.9 1.2
1.5D ARFI Lateral 8.1 0.25 32.9 2.9
Elevation 17.7 1.0 17.8 13.2