Table 4.

The effect of number of bag actuations on emptying and AS deaggregation with the 3.1–565-ECG device using Case 3 piercing [n=3; Mean(SD)]. The data for four actuations is repeated from Table 1 for comparison.

Description Two actuations Four actuations
ED (%) 72.1 (2.5) 71.1 (1.0)
FPF<5μm/ED (%) 77.6 (8.3) 83.1 (1.4)
FPF<1μm/ED (%) 26.1 (5.4) 30.3 (1.6)
MMAD (μm) 1.48 (0.11) 1.39 (0.05)