Table 8.
The effect of orifice size (3.1 vs. 2.3 mm) and 3D rod configuration on the mean AS emitted dose and aerosolization performance with four bag actuations. The data for the 3.1–565-ECG device is taken from Table 1 for comparison.
Description | 3.1–565-ECG | 2.3–343-ECG | 2.3–212-ECG |
---|---|---|---|
ED (%) | 71.1 (1.0) | 75.3 (2.5) | 68.8 (4.4) |
FPF<5μm/ED (%)* | 83.1 (1.4) | 65.4 (6.4)** | 60.2 (8.5)** |
FPF<1μm/ED (%)* | 30.3 (1.6) | 20.7 (3.0)** | 21.1 (4.7)** |
MMAD (μm)* | 1.39 (0.05) | 1.64 (0.07)** | 1.59 (0.07)** |
P<0.05 significant effect of device on % FPF<5μm/ED, FPF<1μm/ED and MMAD (one-way ANOVA).
P<0.05 significant differences compared to 3.1–565-ECG (post-hoc Tukey).