Skip to main content
. 2010 Mar 10;22(1):1–5. doi: 10.1016/j.jsha.2010.03.001

Table 2.

Discrepancy between Datascope Accutorr Plus™ (device) and mercury sphygmomanometer (observer) measurement for systolic blood pressure.

Absolute difference between device and observer (mmHg) Percentage of difference between the device and observer Cumulative percent
⩽5 66.3 66.3
⩽10 27.5 93.8
⩽15 6.3 100