Table 2.
Discrepancy between Datascope Accutorr Plus™ (device) and mercury sphygmomanometer (observer) measurement for systolic blood pressure.
Absolute difference between device and observer (mmHg) | Percentage of difference between the device and observer | Cumulative percent |
---|---|---|
⩽5 | 66.3 | 66.3 |
⩽10 | 27.5 | 93.8 |
⩽15 | 6.3 | 100 |