Table 1.

Validation results for DTNB and DLM methods.

Validation test Acceptance criteria Method result
DTNB DyLight 488
Precision (repeatability) <20% RSD: 2.52%
(denaturing-reducing conditions).
RSD: 17.3% (native conditions).
RSD: NT%
(denaturing-reducing conditions).
RSD: 13.3% (native conditions).

Accuracy Total recovery between 95 and 105% in respect to theoretical value. 99.5% recovery using Somatropin (Cys/Protein ratio 3.98) 101.0% recovery using HSA (Cys/Protein ratio 1.01)

Intermediate precision RSD < 10% RSD: 3.8% RSD: 0.66%

Linearity r 2 ≥ 0.980
Ordinate = 0
r 2 = 0.999 native and denaturing-reducing conditions.
Slope = 0.01285 μM−1
r 2 = 0.983 denaturing conditions.
Slope = 0.0093 μM−1
Forced through zero.
r 2 = 0.990 native and denaturing conditions.
Slope = 465,834 μM−1
Forced through zero.

Quantification limit Report value 5 μM 0.1 μM

Selectivity Cysteine-free protein has the same signal of placebo. GA: 0.0010
Placebo: 0.0005
PM: 0.0005
Placebo: 0.0005

NT%: not tested.