. 2013 Jul 24;17(4):R161. doi: 10.1186/cc12840

Table 3.

Predictive values of pupil diameter variations for insufficient analgesia (BPS)

  10 mA tetanic stimulation 20 mA tetanic stimulation 40 mA tetanic stimulation
Best cutoffs for pupil diameter variation (%)
1
5
13
Sensitivity
0.69 (0.55 – 0.84)
0.85 (0.74 – 0.96)
0.85 (0.74 – 0.96)
Specificity
0.67 (0.52 – 0.81)
0.74 (0.61 – 0.88)
0.78 (0.65 – 0.91)
Positive predictive value
0.50 (0.35 – 0.66)
0.61 (0.46 – 0.76)
0.65 (0.50 – 0.80)
Negative predictive value
0.82 (0.70 – 0.94)
0.91 (0.82 – 1.00)
0.91 (0.83 – 1.00)
Youden index 0.36 0.59 0.62

All data except Youden index were 95% confidence interval. BPS Behavioral Pain Scale.