Table 1. Technical specifications for microEEG and Nicolet Monitor.

Product microEEG Nicolet Monitor
Company Bio-Signal Group Corp. CareFusion Corp.
A/D Converter resolution 16 16
Voltage resolution (μV) 0.15 0.15
Maximum input range 10 mV p. to p. 10 mV p. to p.
Sampling rate 1,000 Hz Up to 2,000 Hz
Bandwidth 0.15 – 500 Hz 0.16 – 1,000 Hz
Input Impedance (MΩ) >100 >100
Separately test GND/REF impedance Yes No
Number of Channels 32 64
Channel crosstalk < -126dB < -40 dB
Noise < 2.4 μV p. to p. (0.1 -100 Hz) < 1.5 μV p. to p. (0.1 -70 Hz)