Table 2.

LC/MS accurate mass determination for endoxifen and impurities.

Peak no. Retention time (min)a Area% of major component Sample conc. Retention time (min)b Exact mass of [M + H] + Ion (amu) Formula assignment of neutral species Mass error of assigned formula (ppm)
1 8.940 0.12 0.165 mg/mL 8.940 360.1970 C24H25NO2 −3.32
2 9.400 2.28 0.0165 mg/mL 9.393 374.2118 C25H27NO2 −0.75
3 10.540 100 0.165 μg/mL 10.533 374.2123 C25H27NO2 −2.11
a

In 0.165 mg/mL sample, UV detection.

b

For sample used for accurate mass determination, UV detection.