Table 5.
Results of PET spatial resolution measurements with a point source on the Ingenuity PET–MRI system compared to the GEMINI TF PET–CT system.
Ingenuity TF PET–MR | GEMINI TF PET–CT (Surti et al 2007) |
|||
---|---|---|---|---|
Spatial resolution | FWHM (mm) | FWTM (mm) | FWHM (mm) | FWTM (mm) |
Transverse (1 cm) | 4.7 ± 0.1 | 9.4 ± 0.2 | 4.8 | 9.7 |
Axial (1 cm) | 4.6 ± 0.1 | 9.5 ± 0.1 | 4.8 | 9.6 |
Radial (10 cm) | 5.0 ± 0.1 | 9.9 ± 0.1 | 5.2 | 10.3 |
Axial (10 cm) | 5.0 ± 0.1 | 9.7 ± 0.1 | 4.8 | 9.6 |
Tangential (10 cm) | 5.3 ± 0.1 | 10.5 ± 0.1 | 5.2 | 10.2 |