Comparison of the ELISA performance in the present microwell array and 96-well plate.

Microwell 96-Well Plate
Reagent consumption (μL) Goat anti-h IgA solution 2 100
Blocking buffer 2.5 200
Antigen/sediment solution 2 100
HRP-conjugated goat anti-h IgA solution 2 100
Substrate solution 2.5 100
Washing buffer 42 2400
Total reagents consumption (mL) 0.053 3
Assay time (min) 25 135
LOD (ng/mL) 0.05 5.19