Table 3.

The mean and standard deviations (SDs) of the trabecular microstructural parameters measured at different object locations in the field of views of cone beam CT (CBCT) systems

CBCT system Object location Tb.Th
Tb.Sp
Tb.N
BV/TV
Mean SD Mean SD Mean SD Mean SD
NewTomâ„¢ 5G Centre 9.87 0.62 14.47 0.82 0.32 0.03 31.69 0.98
Left 10.48 0.85 18.59 1.66 0.29 0.03 29.63 1.56
Right 10.45 0.87 18.02 1.85 0.28 0.03 27.59 1.70
Anterior 10.72 0.86 18.24 1.69 0.29 0.02 30.67 0.64
Posterior 10.23 0.69 17.40 1.51 0.26 0.03 25.94 1.05
Accuitomo 170 Centre 7.62 0.66 15.59 1.51 0.39 0.08 23.52 3.48
Left 7.78 0.82 15.11 0.78 0.32 0.05 21.71 2.04
Right 8.15 0.93 14.32 0.92 0.37 0.05 24.61 2.75
Anterior 8.10 0.80 14.09 0.94 0.35 0.05 26.93 2.75
Posterior 8.27 0.79 14.99 0.84 0.34 0.06 25.51 3.09

The trabecular microstructural parameters: BV/TV, bone volume density; Tb.N, trabecular number; Tb.Sp, trabecular spacing; Tb.Th, trabecular thickness.

NewTom 5G is manufactured by QR Verona, Verona, Italy and Accuitomo 170 by Morita, Kyoto, Japan.