Skip to main content
. Author manuscript; available in PMC: 2014 Jun 24.
Published in final edited form as: J Cataract Refract Surg. 2013 Dec;39(12):1864–1871. doi: 10.1016/j.jcrs.2013.05.048

Table 2.

Repeatability of Optical Coherence Tomography Corneal Pachymetric Variables.

Groups Unit: μm Minimum Minimum-Median S-I SN-IT Ymin

PSD ICC PSD ICC PSD ICC PSD ICC PSD ICC
Normal 1.6 1.00 2.2 0.89 4.2 0.94 3.9 0.93 241 0.85
Keratoconus 3.3 1.00 5.1 0.99 14.2 0.98 12.8 0.97 323 0.84

S-I = superior-inferior, SN-IT = superonasal-inferotemporal, Ymin = vertical location of the minimum.

PSD = pooled standard deviation, ICC = intraclass correlation coefficient.