Skip to main content
. 2014 Jun;58(6):3481–3484. doi: 10.1128/AAC.00025-14

TABLE 3.

Stability testing results for ertapenem

Test conditiona % bias for sample
LOW MED HIGH
Benchtop, RT, 24 h −11.9 −10.7 −14.8
Refrigerator, 4°C, 96 h 5.6 −0.1 −4.3
Freeze-thaw, −20°C, 3 cycles −0.2 0.1 −3.3
Autosampler, 10°C, 24 h −10.0 −2.9 −4.8
a