Quality assessment of identification (LTQ Orbitrap Discovery®, mass resolution FTWM 30,000 at 400 m/z). RT: retention time. Averages of 13 injections in two days at five concentrations (2.5 to 40 ng SPX-1/mL; 0.95 to 190 ng PnTX-G /mL).

NRC CRM SPX-1 NRC CRM PnTX-G
RT (min)
Average %
drift
4.75
2
5.92
1
Fragment 1 Fragment 2 Fragment 3 Precursor Fragment 1 Fragment 2 Fragment 3 Precursor
Exact mass 164.1434 444.3108 674.4415 692.4521 164.1434 458.3265 676.4572 694.4677
Mass Accuracy (Δ ppm)
Average −0.3 0.1 0.4 0.5 −0.4 −0.5 −0.1 0.1
Max. −2.0 −1.6 −0.9 0.7 −1.7 −1.8 −1.4 −1.6
Ion Ratios
% of Frag 1 intensity 100 53 50 94 69 100