Table 2.
Parameters | gCCT | ngCCT |
Scan parameters | ||
Technique | Spiral | Spiral |
kV | 120 | 120 |
mAs | 150 | 30 |
Gantry rotation time (ms) | 330 | 330 |
Individual detector width (mm) | 1.2 | 0.6 |
Number of detectors | 20 × 2 | 32 × 2 |
ECG gating | Yes | No |
Prospective ECG modulation of tube current | Yes | No |
Table feed/s (mm) | 14.4 | 86.4 |
Pitch | 0.2 | 1.5 |
Scan range | Carina-Apex | Chest |
DLP (mGy; mean ± SD) | 116 ± 51 | 84 ± 16a |
Effective dose (mSv; factor = 0.014; mean ± SD) | 2.0 ± 0.9 | 1.2 ± 0.5a |
Effective dose (mSv; factor = 0.014; median, IQR) | 1.61; 1.15-2.0 | 1.13; 1.05-1.20a |
Reconstruction parameters | ||
Effective slice width (mm) | 3.0 | 5.0 |
Reconstruction increment (mm) | 1.5 | 5.0 |
FOV (mm) | 140-160 | 250-300 |
Kernel filtering | B35f1 | B30f(medium-smooth) |
The Table shows the scan parameters for gCCT and ngCCT. gCCT vs ngCCT,
P < 0.05;
Dedicated filter for calcium scoring. gCCT: ECG gated cardiac computed tomography; ngCCT: Non-gated chest computed tomography; FOV: Field of view; ECG: Electrocardiogram; DLP: Dose length product.