Precision and accuracy of the nano-LC/SRM-MS method to determine cT84.66 in mouse plasma

Nominal concentration
(ng/ml)
Calculated concentration
(ng/ml) (mean ± SD)
Intra-day
RSD (%)
Inter-day
RSD (%)
Relative error
(%)
32.3 30.3 ± 4.1 11.2 10.3 −5.9
645 686 ± 52 7.9 13.7 6.4
29025 27516 ± 2537 7.3 10.1 −5.2