Table 1.

Typical imaging parameters used to acquire different CMR images at 1.5T and 3T

Imaging Method Cine Native T1 map Native T2 map LGE
Field Strength 3T 1.5T 3T 1.5T 3T 1.5T 3T 1.5T
Sequence Balanced SSFP Modified Look-Locker Inversion Recovery T2 – prepared SSFP IR – prepared GRE
TR/TE (ms) 3.2/1.6 3.5/1.75 2.2/1.1 2.4/1.2 2.8/1.4 2.2/1.1 3.0/1.5 3.5/1.75
Flip Angle 50° 70° 35° 50° 70° 25° 40°
Bandwidth (Hz/pixel) 1371 930 1042 1002 1371 1002 586 1002
In-plane Resolution 1.3 × 1.3 mm2
Slice Thickness 6mm
Other Parameters 25–30 cardiac phases 8 TIs; 2 inversion blocks of 3+5 images; minimum TI = 110ms; TI increment = 80ms T2-preparation times of 0, 24 and 55ms Optimal TI to null the remote myocardium