Operative Parameters

Study Anesthesia Real-Time Imaging Wavelength (Laser Source)a Number of Fibers Energy (Joules), Power (Watt), Time (Sec) Cancer Volume (cm3) Location Thermal Damage Monitoring Visible Dimension of Thermal Necrosis
Amin Z et al34 Sedation and local anesthesia Transrectal ultrasound 805 nm (Diomed diode) 3 3000 J, 2 W, 500 s Unknown Peripheral zone No Unknown
Atri M et al35 General Transrectal CEUS 830 nm (Indigo diode) 2 2880 J, 2–15W (temperature control at 100°;C), 720 sec 0.25 Mid-posterior region Yes 176
Lindner U et al35 General Transrectal CEUS Unknown 1 or 2 Unknown Mean 0.25 (range, 0.1–0.99) Unknown Yes 300–4000
Raz O et al37 General MRI-guidance and CEUS 980 nm (Visualase diode laser) ≥2 Unknown Unknown Medial mid-prostate; peripheral zone Yes Unknown
Lindner U et al19 General CEUS Unknown 2 or 3 3260–5900 J Not reported Not reported Yes 2500–4500
Woodrum et al39 General MRI-guidance Unknown 2 or 3 Unknown 2.56 Right of urethral anasto-mosis at the bladder base anterior to the rectum Yes Unknown
Lindner G et al38 General Robotic MRI-guidance Unknown Unknown Unknown 0.79; 0.24 Both cases irregularly shaped lesions, obliquely angled Yes 8700–9300
Oto A et al16 Sedation and local anesthesia MRI-guidance 980 nm (Visualase diode laser) 1 2967 J, 11.5 W, 258 s Max diameter 7 mm (range, 4–12) All in peripheral zone Yes Unknown

aDiomed, Andover, MA; Indigo® manufactured by Ethicon Endo-Surgery, Cincinnati, OH; Visualase, Houston, TX. CEUS, contrast-enhanced ultrasound; MRI, magnetic resonance imaging; US, ultrasound.