Published in final edited form as: Phys Med Biol. 2014 Mar 12;59(7):1679–1700. doi: 10.1088/0031-9155/59/7/1679

Table 5.

Measured and simulated temperatures at the end of the sonication for each patient. Sonication index for patient A in parentheses. For the measurements, thermal data from a 3 × 3 voxel neighbourhood centered at the focus is shown: average temperature at the end of the sonication with standard deviation, and in parentheses minimum and maximum temperature. For the simulations, the peak focal temperature after MR-voxel averaging is shown, as well as the focal pressure amplitude.

Patient Measured (°C) Simulated (°C) Focal pressure (MPa)
A (1) 53.1 ± 2.4 (48.8/56.3) 51.5 3.87
A (2) 45.2 ± 1.3 (43.3/47.2) 44.2 2.65
A (3) 53.0 ± 1.7 (50.8/55.7) 46.3 2.75
A (4) 51.4 ± 3.4 (44.6/54.9) 47.5 2.94
B 54.1 ± 2.5 (50.5/57.8) 48.6 3.60
C 56.8 ± 2.5 (53.7/60.9) 51.2 3.45
D 60.5 ± 1.6 (58.4/62.8) 58.8 5.24
E 52.7 ± 2.5 (49.0/57.2) 47.3 3.16