. 2006 Jul 14;12(26):4199–4202. doi: 10.3748/wjg.v12.i26.4199

Table 3.

Results of bacterial contamination after gastroscope reprocessing in both groups

Enzymatic detergent (n = 130) Chlorhexidine (n = 130) P
Type of endoscope (Olympus:Pentax) 60:70 60:70
Positive culture (> 200 cfu/mL) 6 (4.6%) 4 (3.1%) 0.747a
Single organism 5 (3.8%) 1 (0.8%) 0.213b
Mixed organism 1 (0.8%) 3 (2.3%) 0.622b
Pseudomonas aeruginosa 4 (3.1%) 5 (3.8%) 1.000b
Non Pseudomonas spp. 3 (2.3%) 3 (2.3%) 1.000b

a: chi square test; b: Fisher’s exact test.