Table 4.

Comparison of diagnostic accuracy of FibroScan and other markers

Fibrotic Area >10%
OC Sens Speci SE 95% CI AUC
FibroScan 9.1 100 76.9 0.047 0.777 - 0.999 0.932
HA 96.0 76.5 84.6 0.080 0.618 - 0.924 0.803
Col 6.6 52.9 100 0.081 0.615 - 0.923 0.801
APRI 0.57 76.5 84.6 0.081 0.615 - 0.923 0.801
Fibrotic Area > 20%
OC Sens Speci SE 95% CI AUC
FibroScan 13.6 100 95.5 0.024 0.868 - 1.000 0.991
HA 106.0 100 77.3 0.061 0.797 - 0.993 0.946
Col 6.6 87.5 90.9 0.071 0.758 - 0.985 0.918
APRI 0.58 87.5 68.2 0.098 0.642 - 0.937 0.824

OC: Optimal cut off level; Sens: Sensitivity; Speci: Specificity; SE: Standard error; CI: Confidence interval; AUC: Area under the curve; HA: Hyaluronate; Col: Type IV collagen; APRI: AST-to platelet ratio index.