Table 23. Microbiological quality standards of different regulatory agencies.

  ANSI/AAMI/ISO763, 764, 765 ERA-EDTA guidelines765a
Water for dialysis
 Bacteria (CFU/ml) <100 (action level at 50) <100
 Endotoxin (EU/ml) <0.5 <0.25
     
Dialysate
 Bacteria (CFU/ml) <100 (action level at 50) <100
 Endotoxin (EU/ml) <0.5 <0.25
     
Ultrapure dialysate
 Bacteria (CFU/ml) <0.1 <0.1
 Endotoxin (EU/ml) <0.03 <0.03
     
Substitution fluid for infusion
 Bacteria (CFU/ml) Sterile <10−6
 Endotoxin (EU/ml) Undetectable <0.03

AAMI, Association for the Advancement of Medical Instrumentation; ANSI, American National Standards Institute; CFU, colony-forming units; ERA-EDTA, European Renal Association—European Dialysis and Transplant Asssociation; EU, endotoxin units; ISO, International Organization for Standardization.