TABLE IV.

Functional Analysis of Variance Test Results for Intersystem Difference for Low and High Frequency Ranges.

Tumor Bandwidth
(MHz)
Number of
systems
compared
F-statistic Bootstrap
p-value
Effect size
(RSNR)
R1 3.9 to 4.9 8 3.24 0.004 0.669
R1 6.55 to 7.9 4 23.0 <0.0005 2.10
R2 3.9 to 4.9 7 11.3 <0.0005 1.44
R2 6.55 to 7.9 4 13.7 <0.0005 1.59
R3 3.9 to 4.9 8 19.9 <0.0005 1.94
R3 6.55 to 7.9 5 20.7 <0.0005 1.98
R4 3.9 to 4.9 7 19.6 <0.0005 1.93
R4 6.55 to 7.9 5 0.661 0.694 0