Skip to main content
. Author manuscript; available in PMC: 2014 Jul 14.
Published in final edited form as: Biomed Tech (Berl). 2013 Dec;58(6):565–576. doi: 10.1515/bmt-2012-0062

Figure 2. Imaging FFL trajectory for a single 2D projection scan.

Figure 2

Each projection image is produced through the application of slowly varying fields in x and z (in scanner coordinates), and a rapidly varying drive field in z. The signal is received only along the z-axis. 2D projection imaging does not require mechanical movement of the sample; however, the FOV can be optionally extended in the z-direction with a translation stage. Here, a sparse trajectory is shown for illustration purposes, while the actual scanning trajectory is considerably denser.