Published in final edited form as: Ann Thorac Surg. 2014 May 22;98(1):97–102. doi: 10.1016/j.athoracsur.2014.04.027

Table 1.

Mean laboratory values for six sheep supported on the BioVAD circulatory support device. P value determined by ANOVA.

Laboratory Value Baseline Day 1 Day 2 Day 3 Day 4 Day 5 P value
Plasma free hemoglobin (mg/dL)a 2.1 ± 1.6 4.4 ± 4.1 3.5 ± 2.7 2.4 ± 1.8 2.6 ± 1.9 1.2 ± 0.4 0.002
Platelets (K/mm3)b 223 ± 143 142 ± 85 108 ± 64 128 ± 72 182 ± 179 232 ± 245 0.05
Lactic acid (mmol/L)c 1.8 ± 0.9 1.2 ± 0.8 0.7 ± 0.3 0.5 ± 0.2 0.5 ± 0.2 0.5 ± 0.3 0.002
Creatinine (mg/dL)d 1.0 ± 0.2 1.0 ± 0.3 0.9 ± 0.6 0.7 ± 0.1 0.7 ± 0.1 0.7 ± 0.1 <0.001
Total bilirubin (mg/dL)e 0.1 ± 0.08 0.3 ± 0.16 0.2 ± 0.15 0.2 ± 0.11 0.2 ± 0.08 0.2 ± 0.08 0.46
Hemoglobin (g/dL)f 7.2 ± 0.7 6.9 ± 1.1 6.6 ± 0.7 6.7 ± 0.7 6.7 ± 0.8 6.3 ± 0.8 0.91

Reference ranges:

a

1.0-8.0 m/dL;

b

250-750 K/mm3;

c

0.5-2.2 mmol/L;

d

1.2-1.9 mg/dL;

e

0.1-0.5 mg/dL;

f

9.0-15.0 mg/dL