Published in final edited form as: IEEE Trans Ultrason Ferroelectr Freq Control. 2013 Dec;60(12):2471–2487. doi: 10.1109/TUFFC.2013.2847

Table 6.

Comparison of ARF beam sequence sensitivity and specificity for plaque detection and characterization (measured ex vivo in porcine femoral arteries under simulated in vivo conditions) versus published sensitivity and specificity for MRI and CT (measured in vivo in human carotids) [40, 41].

ARF-Based Ultrasound CT MRI

(Ex vivo, procine femoral) (In vivo, human carotid)

Detection Sensitivity Specificity Sensitivity Specificity Sensitivity Specificity
Type III-Va,b,c 85% 79% - - 80–84% 90–98%
Type unreported - - 88–95% 90–96% - -

Characterization

Calcium 96%* 85%* 100% 100% 76–80% 86–94%
Lipid/soft 80%* 86%* 76% 74% 91–98% 65–100%
Fibrous cap 86%* 82%* - - 81% 90%
Degraded IEL 92% 75%* - - - -
Collagen 78% 77% 86% 94% - -

Values marked with ‘*’are reported from SP3-SRx sequence results

Values marked with ‘†’ are reported from SP1.5-ParRx sequence results.