Table 3.

Validity of Horiba analyzer: CENTRIFUGE HORIBA versus CENTRIFUGE HPLC for sweat [Na+] and [K+].

Sweat [Na+] Sweat [K+]
Mean difference ± SD (mEq/L) 4.71 ± 7.87* 0.44 ± 0.52*
95% CI of mean difference (mEq/L) 3.64–5.78 0.34–0.54
ICC 0.96* 0.87*
SEE (mEq/L) 5.56 0.39
TEM (mEq/L) 5.56 0.37
CV (%) 9.37 8.36

n = 210 for sweat [Na+] and 116 for sweat [K+]. CV. coefficient of variation; CI, confidence interval; HPLC, ion chromatography using the Dionex ICS‐3000; HORIBA, Horiba B‐722 for sweat [Na+] and Horiba B‐731 for sweat [K+]; ICC, intraclass correlation coefficient (based on two‐way mixed ANOVA, absolute agreement, average measures); SD, standard deviation; SEE, standard error of the estimate; TEM, typical error of measurement; *P < 0.001.