Table 5.

Validity of field technique using syringe extraction of sweat and compact Horiba analysis: SYRINGE HORIBA versus CENTRIFUGE HPLC for sweat [Na+] and [K+].

Sweat [Na+] Sweat [K+]
Mean difference ± SD (mEq/L) 3.97 ± 10.87* 0.50 ± 0.48*
95% CI of mean difference (mEq/L) 2.49–5.45 0.42–0.59
ICC 0.93* 0.84*
SEE (mEq/L) 8.68 0.44
TEM (mEq/L) 7.68 0.34
CV (%) 12.54 8.48

n = 210 for sweat [Na+] and 116 for sweat [K+]. CV. coefficient of variation; CI, confidence interval; HPLC, ion chromatography using the Dionex ICS‐3000; HORIBA, Horiba B‐722 for sweat [Na+] and Horiba B‐731 for sweat [K+]; ICC, intraclass correlation coefficient (based on two‐way mixed ANOVA, absolute agreement, average measures); SD, standard deviation; SEE, standard error of the estimate; TEM, typical error of measurement; *P < 0.001.