Table 2.
Dwell time deviation for Nucletron RALS measured with a WIC
Dwell time (s) | 0.1 | 0.2 | 0.5 | 1.0 | 5.0 | 15.0 | 30.0 |
Deviation (ms) | 18 | 17 | 23 | 23 | 29 | 23 | 32 |
Dwell time deviation for Nucletron RALS measured with a WIC
Dwell time (s) | 0.1 | 0.2 | 0.5 | 1.0 | 5.0 | 15.0 | 30.0 |
Deviation (ms) | 18 | 17 | 23 | 23 | 29 | 23 | 32 |