T2 relaxation times (mean ± SD) measured with LASER, CP-LASER and T2ρ-LASER sequences and T2 ratios for CP-LASER and LASER and T2ρ-LASER and LASER obtained in the rat brain (n = 6) in vivo at 9.4 T. T2s of all measured metabolites were significantly longer (P < 0.05) with CP pulse sequences compared to LASER. T2 of water was also significantly longer with CP-LASER (P < 0.01) and T2ρ-LASER (P < 0.001) compared to LASER. Between CP-LASER and T2ρ-LASER sequences, the T2s of metabolites were not significantly different (P > 0.05) except in the case of water (P = 0.04). The high SDs observed for tCho and mIns were due to their T2 being longer than the maximum TE used in this study. The goodness of fit, R2 was > 0.9 for all metabolites except for the multiplet of NAA which was 0.88 with LASER.