DCA target accuracy and insertion results.

No. Target accuracy (mm) Insertion time (min) Intracochlear contacts Angular insertion depth (°) Implanted scala
1L 0.79∗ 5 8 of 12 270 SV
1R 0.60∗ 5 10 of 12 330 ST
2L 0.07 5 7 of 12 210 ST
2R 0.15 2 8 of 12 300 ST
3L 0.05 5 12 of 12 420 ST
3R 0.33∗ 2 11 of 12 360 ST
4L 0.28 5 8 of 12 290 ST
4R 0.24 7 9 of 12 360 ST
5L 0.27∗ 5 11 of 12 300 ST
5R 0.22 4 12 of 12 350 ST

Avg. ± SD 0.30 ± 0.23 5 ± 2 10 ± 2 319 ± 58