TABLE II.

WINeR-6 Benchmarking

Reference [31] [2] [32] [4] This Work
Total # of Channels 64 100 128 32 32
Simultaneous Channels 2 1 128 32 32
LNA Gain (dB) 60 60 60 46 67.8/78
Power Source Inductively powered Inductively powered Battery powered Battery powered Inductively powered
Modulation OOK FSK UWB FSK PWM/FSK/OOK
Data Carrier Frequency 174 ~ 216 MHz 915 MHz 4 GHz 3.7 ~ 4.1 GHz 433/915 MHz
Power Carrier Frequency 4 MHz/8 MHz 2.765 MHZ N/A N/A 13.56 MHz
Data Acquisition Spike data reduction Spike data reduction data reduction/raw Raw data Raw data
Data rate 2 Mbps 157 kbps 90 Mbps 24 Mbps 58 ~ 709 kSps
Resolution (bits) 8 10 9 12 8.1
Data link Energy per bit N/A 3185 pJ/b 17.78 pJ/b 1250 pJ/b 607 pJ/b
Power supply (V) 1.8 3.3 3.3 5 3
Process 0.5-µm CMOS 0.6-µm BiCMOS 0.35-µm CMOS Discrete 0.5-µm CMOS
Power consumption (mW) 14.4 8 6 142 5.85