. 2014 Mar 26;33(3):191–199. doi: 10.14366/usg.14008

Table 2.

Comparison of the ablation size estimated on the planning ultrasonograms and specimens

Mode Pig ID Size in the ultrasonograms (A) Size on gross pathology (B)


Length (mm) Width (mm) Length (mm) Width (mm)
Single-point sonication (80 W) 1 8.0 7.4 10.7 7.3
2 7.0 7.3 12.4 5.5
3 7.0 7.3 13.6 7.7
4 6.0 7.0 10.9 6.4
5 8.0 7.4 11.4 4.9
Single-point sonication (96 W) 1a) - - - -
2 10.5 8.0 9.8 6.2
3 9.5 7.6 11.1 7.3
4 8.5 7.5 10.9 6.5
5 11.0 7.8 12.1 6.0
Volume sonication (72 W) 1 19.0 19.8 28.9 19.6
2 16.5 19.3 18.8 9.5
3 21.0 19.8 17.9 18.1
4 22.0 19.8 22.3 19.4
5 18.5 19.8 21.3 24.0
Volume sonication (80 W) 1 15.4 13.9 14.0 11.4
2 13.8 13.9 19.6 12.6
3 18.2 14.4 20.4 11.3
4 15.4 13.9 21.3 17.8
5 17.2 14.0 22.1 20.0
a)

These data were excluded from analysis because the high-intensity focused ultrasound was terminated early, based on the results of the acoustic cavitation monitoring.