TABLE I.
Description | Information sources | Fluoroscopy | Acquisition | Additional information |
---|---|---|---|---|
Method 1 | Patient protocol produced by fluoroscope (reference air kerma, kerma area product, kVp, added filtration); DICOMa header (distance source to patient) | Wholesale, all reference air kerma considered as a whole | Piecemeal, each series considered individually | “Standard” method based on Refs. 12 and 13. |
Method 2 | Patient protocol produced by fluoroscope (reference air kerma); DICOM header (distance source to patient) | Wholesale, all reference air kerma considered as a whole | Piecemeal, each series considered individually | Single backscatter factor of 1.4 used; source-to-patient distance equal to patient table height for last acquisition series. |
Method 3 | Patient protocol produced by fluoroscope (reference air kerma) | Wholesale, all reference air kerma considered as a whole | Piecemeal, each series considered individually | Single backscatter factor of 1.4 used; location of interventional reference point used for source-to-patient distance. |
Method 4 | RDSR | Piecemeal, each series considered individually | Piecemeal, each series considered individually | Used RDSR as data source. |
DICOM = Digital imaging and communications in medicine.