Skip to main content
. 2014 Jun 24;41(7):071913. doi: 10.1118/1.4884020

TABLE II.

Sources of uncertainty in Gafchromic film dosimetry and their associated magnitudes.

Uncertainty source Magnitude (%)
Calibration uncertainty (95% CIa) (Ref. 16) ±6.4
Polarization effects (Ref. 25) ±3
Energy dependence (Ref. 16) ±5
Free-in-air calibration used in backscatter conditions (Ref. 16) −8
Scanner nonuniformity (Ref. 16) ±5
Limits on measured peak skin dose in this studyb [−12.8, 10.0]
a

CI = Confidence interval.

b

Quadratic sum of individual uncertainty sources.