Skip to main content
. 2014 Jun 30;41(7):072303. doi: 10.1118/1.4884224

TABLE I.

Mean values and standard deviations (Std) of DSC, sensitivity, PPV, and ASD between automatic segmentations and manual segmentations for our proposed method and three other methods on the internal dataset. NA in the table means the corresponding measurement was not reported in the literature. The best performance of each measurement is shown in bold lettering.

Method Image No. DSC (in%) Sensitivity (in%) PPV (in%) ASD (in mm)
ASM 75 74.5 ± 11.3 75.7 ± 18.3 79.1 ± 11.8 4.16 ± 3.64
Liao et al. (Ref. 21) 30 86.7 ± 2.2 NA NA 1.90 ± 1.60
Liao et al. (Ref. 20) 66 88.3 ± 2.6 NA NA 1.8 ± 0.9
Our method 75 89.1 ± 3.6 89.9 ± 7.0 89.0 ± 6.2 1.67 ± 0.61