Table 3.

Randox DOA-V Biochip Immunoassay synthetic cathinone presumptive positive results for 20,017 screened urine specimens, and presumptive positive rates of synthetic cathinones (N=106). Presumptive positive urine specimens (N=97) confirmed by liquid chromatography high resolution tandem mass spectrometry (LC-HRMS).

Total # of
Specimens
(N=20,017)
Randox
Presumptive
Positive
Rate
SC
Presumptive
Positive
Rate
Total # of
Confirmed
Specimens
(N=97)*
Positivity
Rate
(%)**
Total Presumptive Positive Specimens for Synthetic Cathinones 106 0.53% N/A 4 4.1%
Presumptive Positive Specimens for BSI only 66 0.33% 62.3% 2 3.3%
Presumptive Positive Specimens for BSII only 31 0.15% 29.2% 1 3.6%
Total Presumptive Positive Specimens for BSI and BSII 9 0.04% 8.5% 1 11.1%
*

9 presumptive positive specimens were not available for confirmation

**

Positivity Rate (%) = 100 × (number of positive by LC-HRMS/ number of positive by Randox)

Abbreviations: SC (synthetic cathinones)