Skip to main content
. 2014 May-Jun;40(3):269–278. doi: 10.1590/S1806-37132014000300010

Table 1. - Temperature program used in order to determine selenium levels in serum.

Step Temperature, °C Ramp, °C/s Duration, s
Drying 90 10 10
Drying 120 15 20
Pyrolysis 500 10 20
Pyrolysis 1,100 30 30
Auto zero 1,100 0 6
Atomizationa 2,200 2,000 3
Cleaning 2,300 1,000 3
a

Measurement.