Table 1. - Temperature program used in order to determine selenium levels in serum.
Step | Temperature, °C | Ramp, °C/s | Duration, s |
---|---|---|---|
Drying | 90 | 10 | 10 |
Drying | 120 | 15 | 20 |
Pyrolysis | 500 | 10 | 20 |
Pyrolysis | 1,100 | 30 | 30 |
Auto zero | 1,100 | 0 | 6 |
Atomizationa | 2,200 | 2,000 | 3 |
Cleaning | 2,300 | 1,000 | 3 |
Measurement.