TABLE 2.
d5-C16:0-LPA human BALF intra-day evaluation, freeze-thaw, and autosampler stability qualification results
Intra-Day Evaluation | Freeze-Thaw Stability (3 cycles) | Autosampler Stability | ||||
QC Level | Accuracy | Precision (% RSD) | Accuracy | Precision (% RSD) | Accuracy | Precision (% RSD) |
QC1 | 112 | 5.9 | 90.6 | 3.7 | 105 | 6.7 |
4.8 nM | ||||||
QC2 | 95.6 | 2.6 | 89.6 | 4.7 | 109 | 4.6 |
36.1 nM | ||||||
QC3 | 94.2 | 2.4 | 92.7 | 1.7 | 99.2 | 5.2 |
181 nM |