Skip to main content
. 2006 Dec;10(4):175–200. doi: 10.1177/1084713806296386

Figure 12.

Figure 12.

Cumulative failure percentages for the 4 generations of Nucleus receiver-stimulators, all recipients worldwide as of June 30, 2006, adults and children separately: (A) CI22M, (B) CI24M (all models and after modification), and (C) CI24R and CI24RE. Error bars show the 95% confidence intervals. Note the y-axis for (A) is twice that of (B) and (C).(Image by permission of Cochlear Limited)