Skip to main content
. 2014 Jul 23;41(8):081713. doi: 10.1118/1.4889779

TABLE II.

Tumor localization accuracy for individual patients and their standard deviations are shown here. The last column shows errors for the tumor 3D positions. The mean tumor localization accuracy across the population is shown in the last row of the table.

  LR AP SI 3D
Patient # (mm) (mm) (mm) (mm)
1 0.07 ± 0.06 0.52 ± 0.39 0.36 ± 0.25 0.67 ± 0.42
2 0.04 ± 0.03 0.63 ± 0.71 0.30 ± 0.25 0.75 ± 0.70
3 0.14 ± 0.16 0.26 ± 0.21 0.21 ± 0.10 0.40 ± 0.22
4 0.20 ± 0.25 1.02 ± 0.75 0.73 ± 0.82 1.46 ± 0.88
5 0.04 ± 0.04 0.27 ± 0.27 0.24 ± 0.17 0.42 ± 0.24
6 0.14 ± 0.15 0.25 ± 0.25 0.32 ± 0.29 0.45 ± 0.39
7 0.05 ± 0.03 0.28 ± 0.29 0.63 ± 0.52 0.73 ± 0.54
8 0.08 ± 0.06 0.86 ± 0.70 0.22 ± 0.11 0.94 ± 0.64
Overall 0.09 ± 0.13 0.51 ± 0.56 0.37 ± 0.42 0.73 ± 0.63