TABLE III.

Variations in tumor tracking error with respect to changing gantry angle.

  LR AP SI 3D
Gantry angle (mm) (mm) (mm) (mm)
0 0.30 ± 0.29 0.49 ± 0.41 0.95 ± 1.02 1.19 ± 1.05
30 0.30 ± 0.31 0.25 ± 0.19 1.34 ± 1.50 1.46 ± 1.48
60 0.20 ± 0.14 0.52 ± 0.37 1.17 ± 1.21 1.33 ± 1.23
90 0.34 ± 0.34 0.43 ± 0.29 1.22 ± 1.49 1.43 ± 1.45
120 0.13 ± 0.16 0.19 ± 0.14 1.24 ± 1.29 1.32 ± 1.23
150 0.12 ± 0.13 0.27 ± 0.23 1.15 ± 1.02 1.21 ± 1.02
180 0.14 ± 0.15 0.31 ± 0.31 1.03 ± 0.96 1.11 ± 0.98
210 0.20 ± 0.17 0.23 ± 0.20 1.29 ± 1.37 1.39 ± 1.33
240 0.13 ± 0.10 0.54 ± 0.36 0.85 ± 0.88 1.04 ± 0.92
270 0.33 ± 0.32 0.47 ± 0.36 1.03 ± 1.30 1.26 ± 1.30
300 0.11 ± 0.14 0.22 ± 0.14 1.21 ± 1.25 1.30 ± 1.20
330 0.09 ± 0.07 0.23 ± 0.17 1.16 ± 1.19 1.21 ± 1.17
Overall 0.20 ± 0.22 0.35 ± 0.30 1.14 ± 1.17 1.27 ± 1.16