Published in final edited form as: IEEE Trans Med Imaging. 2013 Jul 25;32(11):2114–2126. doi: 10.1109/TMI.2013.2274777

TABLE I.

Accuracy Results for Registration in the Synthetic Dataset. SSD Is Reported for an Intensity Range of 0–1. Displacement Vector Error Magnitudes Are in the Intensity Ramp Sections of the Translated Blocks Only, Given as Mean ± Standard Deviation. The Sliding Organ Registration Is More Accurate Than the Diffusive Registration

SSD Displacement vector
error magnitudes
Before 15.28 4.00 ± 0.00 mm
Diffusive 1.07 0.77 ± 0.88 mm
Sliding 0.17 0.28 ± 0.35 mm