Skip to main content
. Author manuscript; available in PMC: 2014 Jul 28.
Published in final edited form as: J Am Soc Mass Spectrom. 2014 Jan 28;25(4):563–571. doi: 10.1007/s13361-013-0815-6

Table 1.

Tabulated fragmentation, collection, and CID efficiencies for resonant excitation CID under various amplitudes for (a) methionine enkephalin at 338 kHz and (b) angiotensin I 3+ at 336 kHz.

(a) 4.8V 5.6V 6.4V 7.2V 8V 8.8V 9.6V
Ef .16 .25 .37 .57 .65 .73 .52
Ec .93 .85 .78 .77 .50 .32 .09
ECID .15 .21 .29 .44 .33 .24 .05
(b) 4.8V 5.6V 6.4V 6.72V 7.2V 7.68V 8V 8.32V 8.8V 9.6V 10.4V
Ef 0 0 .10 .17 .34 .56 .81 .95 1.00 1.00 1.00
Ec .91 .88 .93 .93 .92 .94 .99 .88 .76 .68 .65
ECID 0 0 .9 .16 .31 .53 .80 .84 .76 .68 .65