Axial length acquisition

Cataract type and grade (LOCS III) Lenstar® (standard-5 LS) n/N (%) IOLMaster® (composite-5 IM) n/N (%) IOLMaster® (composite-20 IM) n/N (%)
PSC
≤4.2 75/77 (97.2) 71/77 (92.2) 77/77 (100)
>4.2 13/27 (48.3) 16/27 (59.2) 19/27 (70.4)
Cortical cataract
≤4.2 83/92 (90.2) 84/92 (91.3) 89/92 (96.7)
>4.2 5/12 (41.6) 3/12 (25) 7/12 (58.3)
Nuclear color
≤5.0 85/94 (90.4) 84/94 (89.4) 90/94 (95.7)
>5.0 3/10 (30) 3/10 (30) 6/10 (60)
Nuclear opalescence
≤5.0 83/93 (89.3) 83/93 (89.3) 89/93 (95.7)
>5.0 5/11 (45.5) 4/11 (36.4) 7/11 (63.6)

Notes: Lenstar® standard-5, Haag-Streit AG (Koeniz, Switzerland); IOLMaster® composite-5; IOLMaster® composite-20, Carl Zeiss Meditec AG (Jena, Germany).

Abbreviations: LOCS, Lens Opacities Classification III scoring system; PSC, posterior subcapsular cataract; n, number of patients where the device was able to successfully obtain an axial length; N, total number of patients in each of the groups.