. 2014 Jul 22;8:1369–1376. doi: 10.2147/OPTH.S62653

Table 2.

Acquisition failures, Lenstar® and IOLMaster®

Subject number Eye Age Nuclear color* Nuclear opalescence* Cortical* PSC* IOL Master® (composite-5 IM) (AL, mm) IOL Master® (composite-20 IM) (AL, mm) Lenstar® (standard-5 LS) (AL, mm) Immersion ultrasound (AL, mm)
Both IOLMaster® 500 and Lenstar® failed
2 OS 65 5.6 6.2 5.9 5.9 22.62
3 OD 84 6.2 5.9 3.9 5.6 24.77
8 OD 81 3.5 2.9 5.3 4.3 24.61
8 OS 81 3.4 2.9 4.9 5.9 24.26
11 OD 80 4.9 4.6 3.2 5.9 24.94
16 OS 61 6.9 6.8 5.9 5.9 23.39
45 OD 65 6.9 6.9 5.9 5.9 23.05
IOLMaster® 500 failed but Lenstar® succeeded
23 OD 63 3.4 4.4 0.7 5.3 21.47 21.98
IOLMaster® 500 succeeded but Lenstar® failed
7 OS 79 3.6 2.9 4.8 5.9 23.84 23.82
17 OS 83 3.7 3.6 2.8 3.9 23.11
22 OD 80 3.8 4.3 3.9 5.9 22.24 22.22
38 OD 84 5.9 5.8 2.7 4.2 23.31 23.27
48 OD 39 2.2 1.9 1.5 4.9 22.63 22.60
56 OS 65 3.2 2.4 1.8 5.3 26.67
62 OD 78 5.1 4.3 2.9 4.9 24.63 24.55
65 OS 80 6.6 6.9 5.9 5.9 22.25
67 OD 44 2.3 2.1 2.1 5.5 23.32 23.32
IOLMaster® 500 composite-20 and Lenstar® succeeded but IOLMaster® 500 composite-5 failed
14 OS 74 4.1 3.9 0.4 2.9 23.25 23.25
26 OD 81 3.4 3.4 5.2 0.3 23.35 23.38
26 OS 81 3.4 3.2 5.3 0.1 23.13 23.14
31 OD 81 3.3 3.2 2.6 0.9 24.21 24.13
38 OS 84 5.7 5.1 2.8 3.8 23.20 23.17
43 OD 62 5.2 5.7 5.9 5.9 22.03 22.01

Notes:

*

Based on LOCS III classification scheme. Lenstar® standard-5, Haag-Streit AG (Koeniz, Switzerland); IOLMaster® composite-5; IOLMaster® composite-20, Carl Zeiss Meditec AG (Jena, Germany).

Abbreviations: AL, axial length; LOCS, Lens Opacities Classification III scoring system; OD, right eye; OS, left eye; PSC, posterior subcapsular cataract.